Sea-Bird Scientific HydroCAT
For longterm water quality monitoring with proven anti-fouling technology
Field proven sensors measure conductivity, temperature, pressure and optical dissolved oxygen. Depending on the application, the HydroCAT can collect high quality data for up to a year. Excellent bio-fouling protection is provided by US EPA-approved anti-foulant devices, integral pump, and unique internal flow path provides stable measurements throughout a deployment.
- Unattended
- Temperature, Pressure, Dissolved Oxygen, Conductivity
- Ideally suited for extended deployments in remote, biologically rich environments
- SDI-12, RS-232, RS-485 (optional)
- Robust: excellent anti-fouling capability - EPA approved anti-foulant device and pumped internal flow path for maximum biofouling protection
- Accurate: each instrument is factory calibrated in a temperature-controlled bath that operates at 2-4 times the accuracy of the instrument.
- Cost Effective: no in-field calibrations required, common deployment duration of three plus month, reducing field costs
|
|
|
Download PDF
Download PDF
Measured Parameters | |
Conductivity | |
Range | 0 ... 70 mS/cm (0 ... 70,000 µS/cm) |
Accuracy | ± 0.003 mS/cm (3 µS/cm) |
Typical Stability | 0.003 mS/cm (3 µS/cm) per month |
Resolution | 0.0001 mS/cm (0.1 µS/cm) |
Temperature | |
Range | -5 ... 45°C |
Accuracy | ± 0.002°C / ± 0.01°C (over 35°C) |
Typical Stability | 0.0002°C per month |
Resolution | 0.0001°C |
Pressure | |
Range | 0 ... 20 m/ 0 ... 100 m/ 0 ... 350 m |
Accuracy | ± 0.1% of full scale range |
Typical Stability | 0.05% of full scale range |
Resolution | 0.002% of full scale range |
Optical Dissolved Oxygen | |
Range | 120% of surface saturation in all natural waters |
Accuracy | ± 0.1 mg/L (3 µmol/kg) or ± 2% whichever is greater |
Typical Stability | < 0.03 mg/L (1 µmol/kg)/100,000 samples (20°C) |
Resolution | 0.007 mg/L (0.2 µmol/kg) |
Electrical | |
Clock Stability | 5 seconds/month |
Acquisition Time | 2.3 ... 3.2 seconds/sample (see manual) |
Power Supply | |
Internal | 7.8 Amp-hour (nominal) battery pack, 257 Kjoules (derated for calculations) |
Optional External Power | 0.25 Amps at 9 ... 24 VDC |
Power Consumption(all with pressure) * | |
Quiescent | 0.0004 Watts |
CTD-DO Sample Acquisition (excluding pump): | Real-time data enabled - 0.17 Watts No real-time data - 0.155 Watts |
CTD-DO Sample Waiting (not sampling, pump running, excluding pump): | Real-time data enabled - 0.056 Watts if receive line valid, 0.016 Watts if receive line not valid No real-time data - 0.016 Watts |
CTD-DO Between Samples | Real-time data enabled - 0.056 Watts if receive line valid 0.0004 Watts if receive line not valid No real-time data - 0.0004 Watts |
Pump | 0.12 Watts |
Communications | 0.065 Watts |
* Power consumption values are for standard RS-232 interface; for optional interfaces, see corresponding manual. | |
General | |
Housing Material | Plastic |
Depth Rating | 350 m (1148 ft) |
Weight | 3.4 kg (7.5 lbs) in air, 1.5 kg (3.3 lbs) in water |